Table of Contents
Definitions and test conditions
System performance
Instruments with multiple options
Standard (no options)
Standard Instrument Performance
Option 400 (four samplers for TRL)
Also includes these configurations combined with
Options 089 and/or 012
Option 007 (mechanical switch)
Option 007/085 Performance
Option 085 (high power)
Also includes these configurations combined with
Options 089 and/or 012
Option 012
Option 089
Specifications and characteristics
Measurement throughput summary
Analyzer options
Capabilities
Software
Accessories
Agilent 8719D, 8720D, and 8722D
Network Analyzers
Data Sheet
8719D 50 MHz to 13.5 GHz
8720D 50 MHz to 20 GHz
8722D 50 MHz to 40 GHz
8720D
50 MHZ – 20 GHZ
NETWORK ANALYZER
PORT 1 PORT 2
ACTIVE CHANNEL ENTRY
RESPONSE
R CHANNEL
STIMULUS
INSTRUMENT STATE
Page view 0
1 2 3 4 5 6 ... 25 26

Summary of Contents

Page 1 - Network Analyzers

233, 4, 56, 7, 8910111213152224Table of ContentsDefinitions and test conditionsSystem performanceInstruments with multiple optionsStandard (no options

Page 2 - Measurement uncertainty

10Option 089Option 089 adds frequency offset mode, allowingthe receiver to be offset from the source frequency.This allows direct conversion loss meas

Page 3 - System performance

11Description Specification CodeFrequency CharacteristicsRange8719D 0.05 to 13.51 GHz S-18720D 0.05 to 20.05 GHz S-18722D 0.05 to 40 GHz S-1Accuracy (

Page 4

12Full frequency band sweep time (ms)1Number of PointsMeasurement 51 201 401 1601(Stepped mode/Swept mode)Single Band Sweep (10 to 12 GHz)Uncorrected

Page 5 - Agilent 8722D Option 400

13Option 010 time domainWith the time domain option, data from transmis-sion or reflection measurements in the frequencydomain are converted to the ti

Page 6

14Option 085, high power system This option is designed to permit the measurementof high power amplifiers at RF levels up to 20 Watts(+43 dBm), with f

Page 7

15Measurement capabilitiesNumber of measurement channels2; each fully independentParametersS11: Forward reflection (input match)S21: Forward transmiss

Page 8

16Source controlSweep limitsSet start/stop or center/span of the stimulusparameter (frequency, power, time) directlythrough the source control keys an

Page 9 - Option 012

17One-port calibrationUses test set port 1, or port 2, or both to correctfor directivity, frequency response, and sourcematch errors. Requires open, s

Page 10 - Option 089

18Display controlDisplay formatsSingle channel dual channel overlay (both traceson one graticule), dual channel split (each trace onseparate graticule

Page 11 - <–65 dBc C

19System CapabilitiesLimit linesDefine test limit lines that appear on the displayfor go/no go testing. Lines may be any combinationof horizontal, slo

Page 12

2This data sheet provides two types of performanceinformation:• System Specifications• Supplemental CharacteristicsSystem specifications describe the

Page 13 - Analyzer options

20GPIB (remote) programmingInterfaceGPIB interface operates to IEEE 488-1978 and IEC625 standards and IEEE 728-1982 recommendedpractices.AddressingThe

Page 14 - Analyzer options (continued)

21Parallel portThis connector is used with parallel (or Centronicsinterface) peripherals such as printers and plot-ters. It can also be used as a gene

Page 15 - Capabilities

22Agilent 85071B Materials Measurement SoftwareDescriptionThe 85071B software uses broadband S-parametermeasurements to determine the electromagneticp

Page 16 - Capabilities (continued)

23A wide range of accessories support the Agilent8720 family of network analyzers, including cali-bration kits, verification kits, cables and adapters

Page 17 - Calibration Features

24Economy calibration kitsInclude the open circuit, short circuit, and fixedtermination standards but not sliding terminationsor gauges. Gauges can be

Page 18

25Test port return cable specificationsConnector Type Frequency Length2Return Insertion Loss Stabillty1, 2±Phase (Test Port to Device) (GHz) cm (

Page 19 - Test sequences

Test port adapter setsThe Agilent 85130 series test port adapter sets pro-tect the test set port when connecting devices tothe test port. These adapte

Page 20

3Agilent 8719D, 8720D with 3.5 mm test portsStandard, Options 400, 012, 089, or any combination ofthese optionsCalibration kit: Agilent 85052B, 3.5 mm

Page 21 - Environmental Characteristics

4Agilent 8722D with 2.4 mm test portsStandard, Options 400, 012, 089, or any combination of these options.Calibration kit: Agilent 85056A, 2.4 mm with

Page 22 - Software

5Agilent 8719D Option 400Agilent 8720D Option 400Agilent 8722D Option 400(with 3.5 mm test ports using TRL)Includes instruments with Options 012 and/o

Page 23 - Accessories

6Agilent 8719D, 8720D, 8722D Option 007Agilent 8719D, 8720D, 8722D Option 085Includes instruments with Options 012 and/or 089Option 007 replaces the s

Page 24

7Agilent 8719D or 8720D with Option 007 or 085Calibration kit: Agilent 85052B 3.5 mm with sliding loadsCables: Agilent 85131F 3.5 mm flexible cable se

Page 25 - Accessories (continued)

8Agilent 8722D with Option 007 or 085Calibration kit: Agilent 85052A 2.4 mm with sliding loadsCables: Agilent 85131F 2.4 mm flexible cable setIF bandw

Page 26 - Equipment rack systems

9Option 012Option 012 adds RF loops that provide directaccess to the A and B samplers in the port 1 andport 2 receivers. This allows transmission meas