Agilent TechnologiesAgilent 85225F Performance Modeling SystemInstallation and User’s Guide
10 Installation and User’s GuideIn This Guide...This guide provides instruction on installing, verifying, and servicing the system, as well as an int
Verifying System Functionality 3Installation and User’s Guide 101 If you encounter a problem✔ Check the system connections and settings:1 system conne
103 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies4Servicing the SystemTo troubleshoot the system 104To r
104 Installation and User’s Guide4 Servicing the System To troubleshoot the systemMost system problems are caused by faulty cabling or switch configur
Servicing the System 4Installation and User’s Guide 105 To remove or replace a system componentFollow these steps to remove a system component for per
106 Installation and User’s Guide4 Servicing the System Following the replacement of a system component, complete the procedure in “Performing the Sy
Servicing the System 4Installation and User’s Guide 107 To receive additional assistanceIf you would like assistance, visit the online assistance web
108 Installation and User’s Guide4 Servicing the System To package the system for transportFollow these instruction to package the system in a shippin
Servicing the System 4Installation and User’s Guide 109 17 Add tip indicators and appropriate labeling18 Secure crate to packaging base using band str
111 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesAEnhancing Measurement AccuracyTo enhance measurement a
112 Installation and User’s GuideA Enhancing Measurement Accuracy To enhance measurement accuracyFollow these instructions to enhance the accuracy of
Installation and User’s Guide 11D CV Subsystem Functional Verification TestsThis appendix includes an Agilent 4284A precision LCR meter functional ve
Enhancing Measurement Accuracy AInstallation and User’s Guide 113 5 Always perform a measurement calibration on the network analyzer prior to making d
114 Installation and User’s GuideA Enhancing Measurement Accuracy Understanding System Measurement CalibrationMeasurement accuracy is degraded by the
Enhancing Measurement Accuracy AInstallation and User’s Guide 115 Performing a Coaxial System Measurement CalibrationComplete the following steps to p
116 Installation and User’s GuideA Enhancing Measurement Accuracy If you encounter a problem✔ Inspect the connectors on the load, open, short, and thr
Enhancing Measurement Accuracy AInstallation and User’s Guide 117 Periodic System Component CalibrationA complete calibration verifies that the system
119 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesBDC Subsystem Functional Verification TestUnderstanding
120 Installation and User’s GuideB DC Subsystem Functional Verification Test Understanding the DC Subsystem Functional Verification TestUse this proce
DC Subsystem Functional Verification Test BInstallation and User’s Guide 121 Performing the DC Subsystem Functional Verification TestComplete these st
122 Installation and User’s GuideB DC Subsystem Functional Verification Test For Agilent E5260A/70B DC subsystems:If you encounter a problem✔ Refer to
123 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesCRF Subsystem Functional Verification TestUnderstanding
12 Installation and User’s GuideTyp efac eConventionsThis guide uses the following typeface conventions to describe various aspects of a particular h
124 Installation and User’s GuideC RF Subsystem Functional Verification Test Understanding the RF Subsystem Functional Verification TestUse the proced
RF Subsystem Functional Verification Test CInstallation and User’s Guide 125 Performing the RF Subsystem Functional Verification TestIf the performanc
126 Installation and User’s GuideC RF Subsystem Functional Verification Test 5 Test the forward transmission mode for channel 1.a From the Trace drop
RF Subsystem Functional Verification Test CInstallation and User’s Guide 127 If you encounter a problem✔ Check the GPIB cable and connection.✔ Check t
129 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesDCV Subsystem Functional Verification TestUnderstanding
130 Installation and User’s GuideD CV Subsystem Functional Verification Test Understanding the CV Subsystem Functional Verification TestUse the proced
CV Subsystem Functional Verification Test DInstallation and User’s Guide 131 Performing the CV Subsystem Functional Verification TestIf the performanc
132 Installation and User’s GuideD CV Subsystem Functional Verification Test If you encounter a problem✔ Check the GPIB cable and connection.✔ Check t
133 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesE1/f Noise Subsystem Functional Verification TestUnders
134 Installation and User’s GuideE 1/f Noise Subsystem Functional Verification Test Understanding the 1/f Noise Subsystem Functional Verification Test
Installation and User’s Guide 13 Contents1 Introducing the Agilent 85225F Performance Modeling SystemPerformance Modeling System Configuration Overvie
1/f Noise Subsystem Functional Verification Test EInstallation and User’s Guide 135 Performing the 1/f Noise Subsystem Functional Verification TestIf
136 Installation and User’s GuideE 1/f Noise Subsystem Functional Verification Test If you encounter a problem✔ Check the GPIB cable and connection.✔
137 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesFUnderstanding the Bias NetworksFeatures 138Characteris
138 Installation and User’s GuideF Understanding the Bias Networks FeaturesUsing the Agilent 11612V K11 and K21 bias networks, you can simultaneously
Understanding the Bias Networks FInstallation and User’s Guide 139 CharacteristicsTable 20 11612V Option K11/K21 Bias Network CharacteristicsParameter
140 Installation and User’s GuideF Understanding the Bias Networks OperationEach bias network includes two bias tees, one for force and one for sense.
141 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesGNetwork Analyzer Performance Specification SummaryNetw
142 Installation and User’s GuideG Network Analyzer Performance Specification Summary Network Analyzer System PerformanceThe following specifications
Network Analyzer Performance Specification Summary GInstallation and User’s Guide 143 Measurement Port CharacteristicsMeasurement UncertaintyMeasureme
144 Installation and User’s GuideG Network Analyzer Performance Specification Summary Transmission Measurements
14 Installation and User’s Guide 1/f Noise, CV, RF, and DC Measurement System Configuration 45Figure 16. System Block Diagram 46Figure 17. 1/f Noise M
Installation and User’s Guide 145 IndexNumerics11612V. See bias networks2.4 mm (f-to-f) adapter, 69, 942.4 mm (f-to-m) adapter, 69, 81, 83, 85, 942.4
146 Installation and User’s GuideIndex Hhandle kit, rack mount and, 70, 71, 105hard disk requirements, controller, 59hardware setup, IC-CAP, 89, 96IIC
IndexInstallation and User’s Guide 147 schematic diagramsDC/RF signal flow, 32, 40, 41, 54, 55semi-rigid cables, 71shipment reception procedure, 65sit
Installation and User’s Guide 15 Precautions for Avoiding Electrostatic Discharge 79To connect the bias networks 80Agilent 4156C Systems 80Agilent 415
16 Installation and User’s Guide If you encounter a problem 122C RF Subsystem Functional Verification TestUnderstanding the RF Subsystem Functional Ve
17 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies1Introducing the Agilent 85225F Performance Modeling Sys
18 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Performance Modeling System Configuration OverviewThe sta
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 19 RF and DC Measurement System ConfigurationIn conjunction
Notices© Agilent Technologies, Inc. 2005No part of this manual may be reproduced in any form or by any means (including electronic storage and retriev
20 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The RF SubsystemS- parameter device characterization is p
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 21 Agilent E5270B 8-Slot Precision Parametric Measurement Ma
22 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Component IntegrationSystem component integration is perf
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 23 Figure 2 System Components
24 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 3 Front Panel System Connections, with Agilent 4156
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 25 Figure 3 Front Panel Connections with Agilent 4156C
26 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 4 Front Panel System Connections, with Agilent E526
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 27 Figure 4 Front Panel Wiring Diagram with Agilent E5260A o
28 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 5 Rear Panel System Connections with Agilent 4156CC
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 29 Figure 5 Rear Panel Wiring Diagram with Agilent 4156C
Installation and User’s Guide 3Safety and Regulatory InformationWarnings, Cautions, and NotesThis installation and user’s guide utilizes the followin
30 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System CV, RF, and DC M easure ment Sys tem C onfigu rationTabl
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 31 Figure 6 Rear Panel Wiring with Agilent E5260A or E5270B
32 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 7 DC/RF Cabling Diagram - DC and RF ConfigurationS
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 33 CV, RF, and DC Measurement System ConfigurationWith the a
34 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The CV SubsystemThe Agilent 4284A precision LCR meter pro
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 35 Figure 9 System Components
36 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 7 Front Panel System ConnectionsComponent Informati
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 37 Figure 10 Front Panel Wiring Diagram
38 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 8 Rear Panel System ConnectionsComponent Informatio
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 39 Figure 11 Rear Panel Wiring Diagram
4 Installation and User’s GuideSafety Symbols and Instrument MarkingsSymbols and markings in documentation and on instruments alert you to potential
40 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 12 DC/RF Cabling Diagram - DC and RF Configuration
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 41 Figure 13 DC/RF Cabling Diagram - Parametric Configuratio
42 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The Low Leakage Switch MainframeThe Agilent E5250A is use
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 43 Figure 14 Rear Panel Wiring Diagram including Low Leakage
44 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 15 DC/RF Cabling Diagram - Parametric Configuratio
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 45 1/f Noise, CV, RF, and DC Measurement System Configuratio
46 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The 1/f Noise SubsystemThe Agilent 35670A dynamic signal
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 47 Figure 17 1/f Noise Measurement Block Diagram
48 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Component IntegrationSystem component integration is perf
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 49 Figure 18 System Components
Installation and User’s Guide 5Operator Safety RequirementsThe following general safety precautions must be observed during all phases of operation o
50 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 10 Front Panel System ConnectionsComponent Informat
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 51 Figure 19 Front Panel Wiring Diagram
52 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 11 Rear Panel System ConnectionsComponent Informati
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 53 Figure 20 Rear Panel Wiring Diagram
54 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 21 DC/RF Cabling Diagram - DC and RF Configuration
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 55 Figure 22 DC/RF Cabling Diagram - Parametric Configuratio
56 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Instrument Control InterfaceInstrument control interface
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 57 The LAN/GPIB GatewayThe Agilent E5810A LAN/GPIB gateway p
58 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 23 Rear Panel Wiring Diagram for LAN/GPIB Gateway
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 59 The System ControllerA customer- furnished UNIX workstati
6 Installation and User’s Guidesafety standards of design, manufacture, and intended use of the product. Agilent Technologies, Inc. assumes no liabil
60 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The Rack Cabinet The system is housed in a 1.6 meter rack
Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 61 Performance Characteristics and SpecificationsSupplementa
62 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System RF Subsystem Performance SpecificationsThe overall perfor
63 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies2Installing the SystemTo prepare the installation site 6
64 Installation and User’s Guide2 Installing the System To prepare the installation siteFollow these steps to prepare the site for system installation
Installing the System 2Installation and User’s Guide 65 To receive the systemFollow these steps to store, inspect, and confirm the system shipment.To
66 Installation and User’s Guide2 Installing the System To unpack the shipment crate containing the rack cabinetThe racked system is shipped upright s
Installing the System 2Installation and User’s Guide 67 3 Unload the system a Remove the two brace bolts attaching the side brace assembly to the bott
68 Installation and User’s Guide2 Installing the System To verify the shipmentUse Table 18 “System Receiving Checklist and Replaceable Parts” to:• co
Installing the System 2Installation and User’s Guide 69 12 Complete the receiving checklist.a Compare the Bill of Materials to the system components r
Installation and User’s Guide 7Mains powerGround the systemBefore applying powerCAUTIONThe mains cable shall be permanently connected to the premise
70 Installation and User’s Guide2 Installing the System 16494A Option 001 Triaxial cable, 1.5 meter16494A Option 002 Triaxial cable, 3 meter16494B Opt
Installing the System 2Installation and User’s Guide 71 5063-9221 Rack mount kit with handles, 3-EIA5063-9222 Rack mount kit with handles, 4-EIA5063-9
72 Installation and User’s Guide2 Installing the System E5252A 10 x 12 switch matrixE5260A 8-slot high speed parametric measurement mainframeE5260A Op
Installing the System 2Installation and User’s Guide 73 E8364B Option 022 Extended memoryE8364B Option 080 Frequency offsetE8364B Option 081 Reference
74 Installation and User’s Guide2 Installing the System To install the work surfaceIf the system will not be used with a probe station, install the wo
Installing the System 2Installation and User’s Guide 75 3 Attach the work surface to the support rails.a Slide the work surface onto the support rails
76 Installation and User’s Guide2 Installing the System To ensure your safety while using the system This product has been designed and tested in acco
Installing the System 2Installation and User’s Guide 77 4 Be aware of potential shock hazards during floating-ground measurements.a Do not touch any o
78 Installation and User’s Guide2 Installing the System Precautions for Performing Floating-Ground MeasurementsIC- CAP measurements can be performed w
Installing the System 2Installation and User’s Guide 79 Precautions for Avoiding Electrostatic DischargeNever operate the system without taking precau
8 Installation and User’s GuideFuses and breakersBefore cleaning the systemOvercurrent protectionStatement of Compliance and Declaration of Conformit
80 Installation and User’s Guide2 Installing the System To connect the bias networksFollow these steps to connect the bias networks to the system, and
Installing the System 2Installation and User’s Guide 81 3 Connect and route the triaxial cable from 4156C HRSMU3 FORCE connector to the port 2 bias ne
82 Installation and User’s Guide2 Installing the System Agilent 4156C Systems with Agilent 41501B Expander BoxTo connect the bias networksStep Action
Installing the System 2Installation and User’s Guide 83 4 Connect the Agilent E8364B test ports to the bias networks.a Connect one end of the port 1
84 Installation and User’s Guide2 Installing the System Agilent E5260A/70B SystemsTo connect the bias networksStep Action Notes1 For Agilent E5260A/70
Installing the System 2Installation and User’s Guide 85 3 Connect the triaxial cables from the E5260A/70B SMUs to the bias networks.a Connect the tria
86 Installation and User’s Guide2 Installing the System 6 Connect the bias networks to the device under test.a Connect one semi-rigid cable to the RF/
Installing the System 2Installation and User’s Guide 87 To switch on power to the systemNOTEProper system operation is dependent on the sequence in wh
88 Installation and User’s Guide2 Installing the System 13 On the 4156C, press System > [MISCELLANEOUS], highlight the 4156C value in the GPIB ADDR
Installing the System 2Installation and User’s Guide 89 To configure the LAN/GPIB gateway for functional verificationIf your system includes an Agilen
Installation and User’s Guide 9Compliance with German Noise RequirementsThis is to declare that this instrument is in conformance with the German Re
91 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies3Verifying System FunctionalityTo choose a verification
92 Installation and User’s Guide3 Verifying System Functionality To choose a verification processSystem functionality can be verified using several di
Verifying System Functionality 3Installation and User’s Guide 93 2 If you would like to manually check the functionality of the Agilent E8364B PNA Ser
94 Installation and User’s Guide3 Verifying System Functionality Understanding the System Functional Verification TestThe system functional verificati
Verifying System Functionality 3Installation and User’s Guide 95 Performing the System Functional Verification TestComplete the following steps to ver
96 Installation and User’s Guide3 Verifying System Functionality 3 Start the IC-CAP software.*PC version:a Select Start > Programs > IC-CAP 2004
Verifying System Functionality 3Installation and User’s Guide 97 7 Open the attenuator test model in IC-CAP.a From the IC-CAP/Main window menu bar, ch
98 Installation and User’s Guide3 Verifying System Functionality 10 Configure the instrument state for calibration.a Disconnect the attenuator from th
Verifying System Functionality 3Installation and User’s Guide 99 13 Make the DC and RF measurements in IC-CAP.a Reconnect the attenuator.b In the Atte
100 Installation and User’s Guide3 Verifying System Functionality This completes the functional verification procedure.15 For systems with the Agilent