Table 1-3 Corrected system performance with type-N device connectors, 85092C electronic calibration module
Network analyzer: E5070B/E5071B, calibration module: 85092C (type-N, 50 Ω) electronic calibration (ECal) module, calibration: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with < 1°C deviation
from calibration temperature, isolation calibration not omitted
Description Specification (dB)
3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz
Directivity 52 52 47
Source match 45 41 36
Load match 47 44 39
Reflection tracking ±0.040 ±0.060 ±0.070
Transmission tracking ±0.039 ±0.069 ±0.136
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)
Magnitude Phase