Atec Agilent-E5071B User Manual Page 4

  • Download
  • Add to my manuals
  • Print
  • Page
    / 39
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 3
4
Table 1-2 Corrected system performance with type-N device connectors, 85032F calibration kit
Network analyzer: E5070B/E5071B, calibration kit: 85032F (type-N, 50 Ω), calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature = 23°C ±5°C with < 1°C deviation from calibration
temperature, isolation calibration not omitted
Description Specification (dB)
3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz
Directivity 49 40 38
Source match 41 36 35
Load match 49 40 37
Reflection tracking ±0.011 ±0.032 ±0.054
Transmission tracking ±0.016 ±0.062 ±0.088
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)
Magnitude Phase