Atec Tektronix-1502 User Manual

Browse online or download User Manual for Equipment Atec Tektronix-1502. Atec Tektronix-1502 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1

TM 9-4935-601-14-3&PTECHNICAL MANUALOPERATOR, ORGANIZATIONAL,ANDDS/GS MAINTENANCE MANUAL(INCLUDING REPAIR PARTS)FORTEKTRONIX1502 TIME DOMAINREFLEC

Page 2

TM 9-4935-601-14-3&PSPECIFICATIONGeneral InformationThe Tektronix 1502 is a portable Time Domain Reflectometer that uses pulses to test cables and

Page 3

TM 9-4935-601-14-3&PCROSS INDEX-MFR. CODE NUMBER TO MANUFACTURERMfr. Code Manufacturer Address City, State, Zip0000M SONY/TEKTRONIX CORPORATION P

Page 4

TM 9-4935-601-14-3&PCROSS INDEX-MFR. CODE NUMBER TO MANUFACTURERMfr. Code Manufacturer Address City, State, ZipT1063 SCHOOFS INC. P O BOX 67 MORAG

Page 5 - LIST OF ILLUSTRATIONS

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number1

Page 6

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number1

Page 7

TM 9-4935-601-14-3&PFig. &Index Tektronix Serial/Model No. MfrNo, Part No. Eff Dscont Qty 1 2 3 4 5 Name & Description Code Mfr Part Numb

Page 8 - REV A FEB 1980

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number2

Page 9

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number2

Page 10

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number2

Page 11

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo. Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number2

Page 12

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number3

Page 13

TM 9-4935-601-14-3&PSPECIFICATIONThe performance limits in this specification are valid with the following conditions:The instrument must have bee

Page 14

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number3

Page 15

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number3

Page 16

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number3

Page 17

TM 9-4935-601-14-3&PFig. &IndexTektronix Serial/Model No. MfrNo, Part No. Eff DscontQty 1 2 3 4 5 Name & Description CodeMfr Part Number4

Page 18 - REV B FEB 1980

TM 9-1430-601-14-3&PSCHEMATICS, ACCESSORIES, AND FIGURES10/1/(10-2 blank)

Page 19

TM 9-4935-601-14-3&PAPPENDIX AMAINTENANCE ALLOCATION CHARTInformation pertaining to Maintenance Allocation Chart (MAC) will be furnished at a late

Page 20

By Order of the Secretary of the Army:JOHN A. WICKHAM, JR.General, United States ArmyOfficial: Chief of StaffR. L. DILWORTHBrigadier General, United S

Page 21

TM 9-4935-601-14-3&PSCHEMATIC-PULSER/SAMPLER10-3/(10-4 blank)

Page 22

TM 9-4935-601-14-3&PSCHEMATIC-TIME BASE GENERATOR10-5

Page 23

TM 9-4935-601-14-3&PSCHEMATIC-VERT AMPL/SLOW RAMP COMPONENTS10-6

Page 24

TM 9-4935-601-14-3&PTable 1-2. (cont)ELECTRICAL CHARACTERISTICSCharacteristics Performance Requirements Supplemental InformationFEET/DIV ControlAt

Page 25

TM 9-4935-601-14-3&PSCHEMATIC-BLOCK DIAGRAM10-7

Page 26

TM 9-4935-601-14-3&PSCHEMATIC-MAIN BOARD COMPONENTS10-8

Page 27

TM 9-4935-601-14-3&PSCHEMATIC-VERTICAL AMP/SLOW RAMP10-9

Page 28

TM 9-4935-601-14-3&PSCHEMATIC-OUTPUT AMPL & HIGH VOLTAGE COMPONENTS10-10

Page 29

TM 9-4935-601-14-3&PSCHEMATIC-OUTPUT AMPLIFIERS & HIGH VOLTAGE10-11

Page 30

TM 9-4935-601-14-3&PSCHEMTAIC-LOGIC BOARD COMPONENTS10-12

Page 31

TM 9-4935-601-14-3&PSCHEMATIC-LOGIC10-13

Page 32

TM 9-4935-601-14-3&PSCHEMATIC-POWER SUPPLY10-14

Page 33

TM 9-4935-601-14-3&PSCHEMATIC-POWER SUPPLY10-15

Page 34

TM 9-4935-601-14-3&PSCHEMATIC-POWER SUPPLY COMPONENTS10-16

Page 35

TM 9-4935-601-14-3&PTable 1-2. (cont)ELECTRICAL CHARACTERISTICSCharacteristics Performance Requirements Supplemental InformationLine Voltage 117 V

Page 36

TM 9-4935-601-14-3&P10-17/(10-18 blank)

Page 37

TM 9-4935-601-14-3&P10-19/(10-20 blank)

Page 38

TM 9-4935-601-14-3&P1502 TDR CABLE TESTERFigure 1. Cabinet and Access Box10-21

Page 39

TM 9-4935-601-14-3&P1502 TDR CABLE TESTERFigure 2.. Chassis10-22

Page 40

TM 9-4935-601-14-3&PAccessories1502 TDR CABLE TESTER10-23/(10-24 blank)

Page 41

TM 9-4935-601-14-3&PFigure 3. Front and Frame1502 TDR CABLE TESTER10-25

Page 42

TM 9-4935-601-14-3&PFigure4. Battery Pack1502 TDR CABLE TESTER10-26

Page 43

PIN: 060428-000

Page 44

TM 9-4935-601-14-3&PTable 1-2. (cont)ELECTRICAL CHARACTERISTICSCharacteristics Performance Requirements Supplemental InformationWater ResistanceOp

Page 45

TM 9-4935-601-14-3&PTable 1-2. (cont)ELECTRICAL CHARACTERISTICSCharacteristics Performance Requirements Supplemental InformationDepthIncluding Pan

Page 46

TM 9-4935-601-14-3&POPERATING INSTRUCTIONSThis section of the manual contains the basic information required for the operation of the 1502 TimeDom

Page 47

TM 9-4935-601-14-3&PThe 1502 operates from the battery supply for a minimum of 5 hours (including 20 chart recordings). If aTEKTRONIX Y-T Chart R

Page 48

TM 9-4935-601-14-3&PFigure 2-3. 1502 Front Panel Controls.7. GAIN Screwdriver adjust to set the gain 13. DISTANCE Indicates the distance from the

Page 49

TM 9-4935-601-14-3&P16. RECORD Two-position lever switch; pushed up and thenreleased, it initiates X-Y recorder or a chartrecorder.17. AC LINEFUSE

Page 50

TM 9-4935-601-14-3&PWARNINGDANGEROUS VOLTAGEis used to operate this equipmentDEATH ON CONTACTmay result if safety precautions are not observed.Nev

Page 51

TM 9-4935-601-14-3&P9. Change mp/DIV to 50 and adjust the POSITION controls so the top of the incident pulse is on thehorizontal centerline.10. Pr

Page 52

TM 9-4935-601-14-3&PThis setting of the FEET/DIV (METRES/DIV) control ensures that the reflected signal will appear in thedisplay window. Measure

Page 53

TM 9-4935-601-14-3&PFig. 2-7 is a chart for converting reflected pulse amplitude to impedance. Rho is dependent on thecharacteristic impedance,

Page 54

TM 9-4935-601-14-3&PFigure 2-8. Open Cable. Figure 2-9. Shorted Cable.CHECKING CABLES WITH IMPEDANCE OTHER THAN 50 ΩCables with a characteristic

Page 55

TM 9-4935-601-14-3&PFigure 2-10. Crimped Cable. Figure 2-11. Frayed Cable.An X-Y recorder can be used with the 1502 by connecting it to the X-Y

Page 56

TM 9-4935-601-14-3&PNominal C-30B/31B camera settings are:f/stop 5.6-11time 1 secmagnification 1.2 (fixed at 0.5 in C31B)focus as necessaryINSTRUM

Page 57

TM 9-4935-601-14-3&PTHEORY OF OPERATIONThis section describes the circuits of the 1502 using a combination of block and circuit diagrams on pull-o

Page 58

TM 9-4935-601-14-3&PSequential equivalent-time sampling is used to develop a display. Two ramps are generated, a fast rampand a slow ramp. The f

Page 59 - Grounded (Neutral) Blue White

TM 9-4935-601-14-3&PDETAILED CIRCUIT DESCRIPTIONPulserThe Pulser consists of three circuits: a Clock Generator, a Logic Programmer, and a Pulse Ge

Page 60

TM 9-4935-601-14-3&PBias current for tunnel diode CR1703 is provided by R1601 and R1693. Transistors Q1602 and Q1603serve as a switch to direct t

Page 61

TM 9-4935-601-14-3&PThis material is reproduced through the courtesy of Tektronix Corporation.Distribution is limited to use with the PATRIOT miss

Page 62

TM 9-4935-601-14-3&PDuring time slots 1 and 2,Q1597 is saturated, discharging memory capacitor C1596. At the start of timeslot 3, Q1597 is cut of

Page 63

TM 9-4935-601-14-3&PA negative-going signal from the sampler comparator is amplified by Q1553 and the collector output isfed to a fast-rise one-sh

Page 64

TM 9-4935-601-14-3&PFigure 3-5. Functional Block Diagram of Slow Ramp Generator.Integrator. An inverting amplifier circuit with an FET input is

Page 65

TM 9-4935-601-14-3&PSweep Inverter with Horizontal Attenuator (Schematic 2)The Sweep Inverter consists of an operational amplifier U1344; selecta

Page 66

TM 9-4935-601-14-3&PGain Amplifier. The Gain Amplifier consists of U2143 with variable gain provided by R0263. Theincoming voltage is attenuated

Page 67

TM 9-4935-601-14-3&PFigure 3-6. Block Diagram of 1502 Power Supply. CAUTIONIf the instrument is wired for 220 V operation, be sure that the two

Page 68

TM 9-4935-601-14-3&PThe output of the astable multivibrator is connected to the fly-back switching circuitry consisting of Q6349(an emitter follow

Page 69

TM 9-4935-601-14-3&PFigure 3-7. Chart Recorder Logic.Horizontal Correction Network. The Horizontal Correction network compares the output of the

Page 70

TM 9-4935-601-14-3&PQ5226 will conduct only when the Strip Chart Recorder is plugged in. Q5237 and Q5229 act as a dc levelshift amplifier between

Page 71

TM 9-4935-601-14-3&PCALIBRATION PROCEDUREThe first part of this procedure checks the electrical characteristics of the 1502 that appear in theSpec

Page 72

TM 9-4935-601-14-3&PTABLE OF CONTENTSPageLIST OF ILLUSTRATIONS...

Page 73

TM 9-4935-601-14-3&PPerformance Check1. Make sure the battery pack is in place and connect the unit to the proper AC power source.2. Pull the POWE

Page 74

TM 9-4935-601-14-3&PPerformance Check11. Remove the 50 Ω terminator and connect the probe adapter to the CABLE connector (SN B040616 andup). Chan

Page 75

TM 9-4935-601-14-3&PPerformance Check21. Check that the jitter is not greater than 0.1 division (200 ps). See Fig. 4-2.22. Change the MULTIPLIER

Page 76

TM 9-4935-601-14-3&PPerformance CheckFigure 4-4. Falltime.32. Connect the probe adapter to the CABLE connector if the connector is the grounding

Page 77

TM 9-4935-601-14-3&PAdjustment Procedure45. Adjust the DISTANCE dial to locate the reflected pulse on each graticule line. The DISTANCE dialshoul

Page 78

TM 9-4935-601-14-3&PAdjustment Procedure3. Remove the EMI shields from the top and bottom of the unit. Place the battery pack in the rear of the1

Page 79 - REV SEP 1983

TM 9-4935-601-14-3&PAdjustment ProcedureFigure 4-5. Power Supply Test Points And Adjustments.24 (a). Locate the pulse with the ZERO REF control a

Page 80

TM 9-4935-601-14-3&PAdjustment ProcedureFigure 4-6. Main Board Test Points.35. Check-that the trace aligns with the graticule line within 0.1 div

Page 81

TM 9-4935-601-14-3&PAdjustment ProcedureTurn the instrument POWER off. Turn the instrument upside down and remove the shield that covers thepulse

Page 82

TM 9-4935-601-14-3&PAdjustment Procedure(e). Set the FEET/DIV to 5, set the TG 501 to 10 ns and check that when the first time mark is alignedwith

Page 83

TM 9-4935-601-14-3&PLIST OF ILLUSTRATIONSFigureNo. PageFigureNo. Page1502 Time Domain Reflectometer vi 5-8 Front Panel, Identification of Parts fo

Page 84

TM 9-4935-601-14-3&PAdjustment Procedure58. Set the FEET/DIV multiplier to X1.59. Set the incident pulse to a graticule reference line with the ZE

Page 85

TM 9-4935-601-14-3&PAdjustment ProcedureFigure 4-9. Time Marks Aligned (Metric).(b). Check that the chart paper in the recorder has a bold line n

Page 86

TM 9-4935-601-14-3&PAdjustment Procedure(b). Set CABLE DIELECTRIC to SOLID POLY. Locate two time marks on the graticule with theDISTANCE dial.(c)

Page 87

TM 9-4935-601-14-3&PAdjustment ProcedureFigure 4-10. Pulse Display.93. Adjust-the front-panel GAIN screwdriver control for a pulse amplitude of 5

Page 88

TM 9-4935-601-14-3&PAdjustment ProcedureFig. 4-12. Reflected-Fall Time Check.106. Remove the 50 n terminator and checkthat the trace shift is le

Page 89

TM 9-4935-601-14-3&PAdjustment Procedure114. Move the test lead from R2229 toR2326 (see Fig. 4-13). The DVM should read0.400 0.008 volt.115. Mov

Page 90 - A4 TM 9-4935-601-14-3&P

TM 9-4935-601-14-3&PAdjustment ProcedureY OUTPUT10. Move the test leads from the PEN LIFT terminals to the corresponding Y terminals of the X-Y OU

Page 91

TM 9-4935-601-14-3&PMAINTENANCENOTEIt is recommended that the 1502 be serviced at a Tektronix Service Center. Theinstrument should be serviced by

Page 92

TM 9-4935-601-14-3&PVisual InspectionPrior to recalibration, the 1502 should be inspected for such defects as broken connections, damagedcircuit b

Page 93

TM 9-4935-601-14-3&PFigure 5-1. DC Fuse Location.NOTEAll screw fasteners used on circuit boards, on or near a cam switch, should betightened to

Page 94

TM 9-4935-601-14-3&POPERATORS SAFETY SUMMARYThe general safety information in this part of the summary is for both operating and servicing personn

Page 95

TM 9-4935-601-14-3&PFigure 5-2. Transistor and Integrated Circuit Lead Configuration.5-4

Page 96

TM 9-4935-601-14-3&PFigure 5-3. Rear View of The HV Board.Logic Board1. Remove the three pin connectors from the Logic board.2. Remove the four s

Page 97

TM 9-4935-601-14-3&P2. Lift up carefully on the rear of the cavity shield, applying heat to the solder connections to the bncconnector if necessa

Page 98 - 8-1/(8-2 blank)

TM 9-4935-601-14-3&PSEALING PROCEDURESNOTEIn the following assembly steps, the part being discussed is identified by its location in the Mechanica

Page 99

b. Procedure1. Clean mating surfaces of seals with acetone.2. Coat mating surfaces with No. 3144 sealant.3. Install seals in subpanel.6. Installa

Page 100 - TM 9-4935-601-14-3&P

TM 9-4935-601-14-3&PNOTE1. Pressing the retainer should cause some of the sealant to appear outside the retainer area, see Fig. 5-7.Failure of th

Page 101

TM 9-4935-601-14-3&PFigure 5-9. Detail of Plug-In Compartment seal. Figure 5-10. Detail of BNC Connector and Washer.4. Clamp Seal in place.5.

Page 102 - Fig. &

TM 9-4935-601-14-3&P3. Apply bead of No. 3144 sealant to each shaft/seal interface.4. Install pushbuttons in their respective locations.12. Att

Page 103

TM 9-4935-601-14-3&PFigure 5-14. Detail of Fuseholder Caps. Figure 5-16. Alignment of Battery Meter3. Insert the caps is the Fuseholders and p

Page 104

TM 9-4935-601-14-3&P2. Screw the bushing into the resistor assembly.3 Form a bead of No. 3144 sealant around the fillet area of the bushing.4.

Page 105

TM 9-4935-601-14-3&PSERVICE SAFETY SUMMARYFOR QUALIFIED SERVICE PERSONNEL ONLYRefer also to the preceding Operators Safety Summary.Do Not Service

Page 106

TM 9-4935-601-14-3&P20. Install the ZERO REFERENCE control in the Front Panel (see Figs. 5-21 and 5-22)a. RESISTOR, VARIABLE MPL Item 2-27.This

Page 107

TM 9-4935-601-14-3&PFigure 5-24. Application of sealant to RECORDtoggle-switch nut.Figure 5-26. Application of sealant to the GAIN orthe 0-dB SE

Page 108

TM 9-4935-601-14-3&P3. Do not install Main Circuit Board holding screws at this point.24. Install cam switch Inserts (see Fig. 5-28)a. INSERT,

Page 109

TM 9-4935-601-14-3&PFigure 5-30. Lubrication of POSITION control knob. Figure 5-31. Installation of knob on shaft..Figure 5-32. Lubrication of

Page 110

TM 9-4935-601-14-3&PFigure 5-34. Lubrication of FOCUS and INTENSITYcontrol knobs.2. Install the knobs on the shafts of the FOCUS and INTENSITY c

Page 111

TM 9-4935-601-14-3&PFigure 5-37. Detail of front panel jacks. Figure 5-38. Detail of X-Y Module thumbscrewsd. JACK, TIP: RED MPL Item 3-6.e.

Page 112

TM 9-4935-601-14-3&P33. Attach Latch Assemblies (2 ea.) to the Front Cover (See Fig. 5-39)a. COVER, FRONT MPL Item 1-29.b. LATCH ASSY. MPL It

Page 113 - 9-15/(9-16 blank)

TM 9-4935-601-14.3&POPTIONSOption 04 Chart RecorderOption 05 Metric VersionOption 76 P7 PhosphorPOWER CORD OPTIONSOption A1: 220 V/16 A, Universal

Page 114 - 10/1/(10-2 blank)

TM 9-4935-601-14-3&PREPLACEABLEELECTRICAL PARTSPARTS ORDERING INFORMATIONReplacement parts are available from or through your local Tektronix, Inc

Page 115 - A-1/(A-2 blank)

TM 9-4935-601-14-3&PCROSS INDEX-MFR. CODE NUMBER TO MANUFACTURERMfr. CodeManufacturer Address City, State, Zip0000M SONY/TEKTRONIX CORPORATION P O

Page 116

TM 9-4935-601-14-3&P1792-011502 Time Domain Reflectometer.REV A FEB 1980vi

Page 117 - SCHEMATIC-PULSER/SAMPLER

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA1 670-4338-00 CKT BOARD

Page 118 - SCHEMATIC-TIME BASE GENERATOR

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberC1435 283-0003-00 CAP.,FX

Page 119

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberQ1317 151-0302-00 TRANSIS

Page 120 - SCHEMATIC-BLOCK DIAGRAM

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR1148 321-1210-01 RES.,FX

Page 121

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR1509 315-0510-00 RES.,FX

Page 122

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR1651 315-0472-00 RES.,FX

Page 123

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA2 6703009-00 B010100 B04

Page 124

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR2227 321-0614-00 RES.,FX

Page 125

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA3 670-3130-00 B010100 B0

Page 126 - SCHEMATIC-LOGIC

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR3149 315-0224-00 RES.,FX

Page 127 - SCHEMATIC-POWER SUPPLY

TM 9-4935-601-14-3&PSECTION 0.GENERAL0-1. MAINTENANCE FORMS AND RECORDSDepartment of the Army forms and procedures used for equipment maintenance

Page 128

A4 TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA4 670-3010-00 B010100

Page 129

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA5 6703513-00 B010100 B02

Page 130 - 10-17/(10-18 blank)

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberR5161 315-0104-00 RES.,FX

Page 131 - 10-19/(10-20 blank)

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code Mfr Part NumberA6 670-3520-00 B010100 B0

Page 132 - 1502 TDR CABLE TESTER

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code mfr Part NumberL6344 108-0598-00 COIL,RF

Page 133 - Figure 2.. Chassis

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code mfr Part NumberR6451 315-0104-00 RES.,FX

Page 134 - 10-23/(10-24 blank)

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code mfr Part NumberA7 670-3865-00 B010100 B1

Page 135 - Figure 3. Front and Frame

TM 9-4935-601-14-3&PTektronix Serial/Model No. MfrCkt No. Part No. Eff Dscont Name & Description Code mfr Part NumberCHASSISBT0369 016-0595-00

Page 136 - Figure4. Battery Pack

TM 9-4935-601-14-3&PSECTION 8DIAGRAMS AND CIRCUIT BOARD ILLUSTRATIONSSymbolsGraphic symbols and class designation letters are based on ANSI Standa

Page 137 - PIN: 060428-000

TM 9-4935-601-14-3&PSECTION 9REPLACEABLEMECHANICAL PARTSPARTS ORDERING INFORMATIONReplacement parts are available from or through your local Tektr

Comments to this Manuals

No comments