Agilent E8257D PSG
Analog Signal Generator
Data Sheet
The Agilent E8257D is a fully synthesized signal generator with high output
power, low phase noise, and optional ramp sweep capability.
Specifications apply over a 0 to 55 °C range, unless otherwise stated, and apply
after a 45 minute warm-up time. Supplemental characteristics, denoted as
typical, nominal, or measured, provide additional (non-warranted) information
at 25 °C, which may be useful in the application of the product.
Definitions
Specifications (spec): Represents warranted performance for instruments with a
current calibration.
Typical (typ): Represents characteristic performance which is non-warranted.
Describes performance that will be met by a minimum of 80% of all products.
Nominal (nom): Represents characteristic performance which is non-warranted.
Represents the value of a parameter that is most likely to occur; the mean and/or mode
of all measurements of a parameter.
Measured: Represents characteristic performance which is non-warranted. Represents
the value of a parameter measured on an instrument during design verification.
Page view 0
1 2 3 4 5 6 ... 23 24

Summary of Contents

Page 1 - Analog Signal Generator

Agilent E8257D PSG Analog Signal GeneratorData SheetThe Agilent E8257D is a fully synthesized signal generator with high output power, low phase noise

Page 2 - Table of Contents

101. Phase noise specifications are warranted from 15 to 35 °C. 2. Measured at +10 dBm or maximum specified power, whichever is less.Option UNX: Abs

Page 3 - Specifications

11Measured phase noise with E5500 and plotted without spursStandard SSB phase noise (measured)1-20-30-40-50-60-70-80-90-100-110-120-130-140-150-160-17

Page 4 - (Option 007)

12Residual FM(RMS, 50 Hz to 15 kHz bandwidth) CW mode < N x 6 Hz (typ)Option UNX/UNR < N x 4 Hz (typ)Ramp sweep mode < N x 1 kHz (typ)B

Page 5

13 Maximum deviation2 Frequency Maximum deviation 250 kHz to 250 MHz 2 MHz > 250 to 500 MHz 1 MHz > 500 MHz to 1 GHz 2 MHz > 1 GHz

Page 6 - Output power (dBm)

14Depth Linear mode Exponential (log) mode (downward modulation only)Maximum: ALC On: > 90% > 20 dB ALC Off with Power Search2

Page 7 - Errorr (dB)

15Modulation types AM, FM, and FMInput impedance 50 or 600 (nom) switchedHigh/low indicator(100 Hz to 10 MHz BW, Activated when input level e

Page 8 - Spectral purity

161. With ALC off, specs apply after the execution of power search. Specifications apply with Atten Hold Off (default mode for instruments with

Page 9 - from carrier (dBc/Hz)

17 10 MHz to 3.2 GHz Above 3.2 GHzWidth compression ±5 ns (typ) ±5 ns (typ) (RF width relative to video out) Video feed-through1 < 125 mv (ty

Page 10 - from carrier

18Interfaces GPIB (IEEE-488.2,1987) with listen and talk, RS-232, and 10BaseT LAN interface.Control languages SCPI version 1997.0. Completely c

Page 11

191. Storage below –20 °C instrument states may be lost.2. As is the case with all signal generation equipment, phase noise specifications are not w

Page 12

2Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Freq

Page 13 - Phase modulation

20RF output Output impedance 50 Ω (nom)Option 520 Precision APC-3.5 male, or Type-N with Option 1EDOptions 532, 540 and 550 Precision 2.4 mm mal

Page 14 - (typical)

21Stop sweep In/Out Open-collector, TTL-compatible input/output. In ramp sweep operation, provides low level (nominally 0 V) during sweep r

Page 15 - Internal modulation source

22Model/option DescriptionE8257D-520 Frequency range from 250 kHz to 20 GHz E8257D-532 Frequency range from 250 kHz to 31.8 GHz E8257D-540 Frequen

Page 16 - (Option UNW)

23For additional information, visit:www.agilent.com/find/psgFor more information about renting, leasing or financing Agilent’s latest technology, visi

Page 17 - Timebase (nsec)

www.agilent.com/fi nd/emailupdatesGet the latest information on the products and applications you select. www.agilent.com/fi nd/agilentdirectQuickly ch

Page 18 - Remote programming

3Range 1Option 520 250 kHz to 20 GHzOption 532 250 kHz to 31.8 GHzOption 540 250 kHz to 40 GHzOption 550 250 kHz to 50 GHzOption 567 250 kHz to 6

Page 19 - General specifications

4Operating modes • Step sweep of frequency or amplitude or both (start to stop) • List sweep of frequency or amplitude or both (arbitrary list)Swe

Page 20 - Input/Output Descriptions

5Power1 (dBm)Frequency range Standard Option 1EA spec. (typ)Option 520: 250 kHz to 3.2 GHz –20 to +155 –20 to +16 (+19) 250 kHz to

Page 21

Step attenuator1 (Option 1E1) Options 520, 532, and 540 0 dB and 5 dB to 115 dB in 10 dB stepsOptions 550 and 567 0 dB to 90 dB in 10 dB stepsMaxim

Page 22 - Options, Accessories

71. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range, with attenuator hold off (normal operating mode).

Page 23 - Literature

8Output impedance 50 Ω (nom)SWR (internally leveled)250 kHz to 2 GHz < 1.4:1 (typ)> 2 GHz to 20 GHz < 1.6:1 (typ)> 20 GHz to 40 GHz

Page 24 - Revised: March 27, 2008

91. Sub-harmonics are defined as Carrier Freq / N). Specifications are typical for sub-harmonics beyond specified frequency range (beyond 50 GH