Atec Agilent-N5173B User Manual

Browse online or download User Manual for Equipment Atec Agilent-N5173B. Atec Agilent-N5173B User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - N5173B Microwave Analog

EXG X-Series Signal GeneratorN5173B Microwave Analog 9 kHz to 13, 20, 31.8, or 40 GHzData Sheet

Page 2 - Definitions

10Option 520 harmonics at +10 dBm (measured)Frequency (MHz)–30–40–50–60–70–80–900 5000 10000 1500020000 250002nd harmonic (dBc)Option 540 harmonics at

Page 3 - Frequency Specifications

11Nonharmonics (CW mode) 1, 2Range > 10 kHz offsetStandard (dBc)9 kHz to < 5 MHz –655 to < 250 MHz –75250 to < 750 MHz –78750 MHz to <

Page 4 - Frequency reference

12Analog Modulation SpecificationsFrequency bandsBand # Frequency range N1 9 kHz to < 5 MHz Digital synthesis2 5 to < 250 MHz 13 250 to < 375

Page 5 - Amplitude Specifications

13Amplitude modulation (Option UNT) 1Depth Linear mode Exponential modeSettable depth ALC ON with deep AM (default) or ALC off 20 to 100% 0 to 50 dBDe

Page 6 - (ALC on) () = typical

14External modulation inputs(Option UNT required for FM, AM, and phase modulation inputs; Option UNW required for pulse modulation inputs)EXT1 AM, FM,

Page 7 - Sweep modes

15Video feed-through 1 < 3.2 / > 3.2GHz < 50 mV (typ) / < 3 mV (typ) Video delay (external input to video) 40 ns, nominalRF delay (video t

Page 8 - () = measured

16General CharacteristicsRemote programmingInterfaces GPIB IEEE-488.2, 1987 with listen and talkLAN 1000BaseT LAN interface, LXI Class C compliantUSB

Page 9

17MemoryMemory is shared by instrument states, user data files, sweep list files, and other files. Option 006 instrument security allows storage of up to

Page 10

18Inputs and OutputsFront panel connectors (all connectors are BNC unless otherwise stated)RF output Output impedance 50 Ω (nom)Option 513/520 Precisi

Page 11 - (measured)

19 USB Type-A There are two USB 2.0 Type-A connectors on the rear panel. Used with a memory stick for transferring instrument states, licenses and ot

Page 12 - Frequency bands

2Table of ContentsDefinitionsDefinitions ...2Frequency specifications ...

Page 13

Agilent X-Series Signal GeneratorsMXG Microwave Signal Generator Data Sheet 5991-3131ENMicrowave Signal Generator Flyer 5991-3594ENX-Series Signal Gen

Page 14 - External modulation inputs

3Frequency SpecificationsRangeFrequency range Option 513 9 kHz to 13 GHzOption 520 9 kHz to 20 GHzOption 532 9 kHz to 31.8 GHzOption 540 9 kHz to 40 G

Page 15

4Frequency referenceAccuracy± aging rate± temperature effects ± line voltage effects± initial setting accuracyInternal time base reference oscillator

Page 16 - General Characteristics

5Output parametersSettable range(with Option 1E1 and 1EA)+30 to –130 dBmSettable range(without Option 1E1 and 1EA)+19 to –20 dBmResolution 0.01 dB, (n

Page 17

6-1-0.8-0.6-0.4-0.200.20.40.60.8102 468101214161820meanmean - std devmean + std devFrequency (GHz) vs. level accuracy (dB) Measured level accuracy at

Page 18 - Inputs and Outputs

7SWR (measured CW mode)Frequency Attenuator state0 dB 5 dB and greater≤ 2 GHz < 1.7:1 < 1.2:1> 2 to 8 GHz < 1.4:1 < 1.4:1> 8 to 13 G

Page 19

8Standard absolute SSB phase noise (dBc/Hz) (CW) [at 20 kHz offset] 1 () = measured5 to < 250 MHz –115 (–120)250 MHz –129 (–134)500 MHz –124 (–128

Page 20 - Related Literature

9Broadband noise 1 () = measured100 MHz (–143 dBc/Hz)500 MHz (–155 dBc/Hz )1 GHz (–163 dBc/Hz )10 GHz (–150 dBc/Hz)20 GHz (–143 dBc/Hz)40 GHz (–135 d