Maximum Acquisition
Points/Ch
( 1 Ch) / (2 Ch) / ( 3 – 4 Ch)
Standard 1M / 500k / 250k 1M / 500k / 250k 1M / 500k / 250k
M – Memory Option 4M / 2M / 1M 4M / 2M / 1M 4M / 2M / 1M
L – Memory Option 16M / 8M / 4M 16M / 8M / 4M 16M / 8M / 4M
VL – Memory Option 32M / 16M / 8M 32M / 16M / 8M 32M / 16M / 8M
XL – Memory Option 64M / 32M / 16M – –
Acquisition Modes
Random Interleaved
Sampling (RIS)
50 GS/s for repetitive signals: 200 ps/div – 1 µs/div
Single-Shot For transient and repetitive signals: 200 ps/div – 1000 s/div
Sequence 2 – 8000 segments
Intersegment Time Typically 30 µs
Acquisition Processing
Averaging
Summed averaging to 103 sweeps (standard).
Continuous averaging up to 106 sweeps with weighting range from 1:1 to 1:1023 (option).
Enhanced Resolution (ERES) From 8.5 to 11 bits vertical resolution
Envelope (Extrema) Envelope, floor, roof for up to 106 sweeps
Triggering System
Modes Normal, Auto, Single, and Stop
Sources
Any input channel, external, Ext/5 or line; slope, level, and coupling unique to each source (except line
trigger)
Slope Positive, Negative, Window
Coupling modes DC, AC, HF, HFREJ, LFREJ
AC Cutoff
Frequency
7.5 Hz Typical
HFREJ, LFREJ
50 kHz typical
Pre-trigger delay 0 – 100% of horizontal time scale
Post-trigger delay 0 – 10000 divisions
Hold-off by time or events Up to 20s or from 1 to 99 999 999 events
Internal trigger range ±5 div
Max trigger frequency 1 GHz (DC, AC), >1 GHz (HF) on WavePro 950, >2.0 GHz (HF) on WavePro 960
External trigger input range ±0.5 ( ±2.5 V with Ext/5 selected )
Maximum ext. input @ 50
Ohms
±5 V DC or 5Vrms
Comments to this Manuals