Frequency range DC coupled AC coupled
Option 503 NA 9 kHz to 3.0 GHz
Option 507 NA 9 kHz to 7.5 GHz
Option 513 9 kHz to 13.6 GHz 10 MHz to 13.6 GHz
Option 526 9 kHz to 26.5 GHz 10 MHz to 26.5 GHz
Band LO multiple (N) AC coupled
RF (Option 503, 507) 0 1 9 kHz to 3.0 GHz
1 1 2.95 to 3.80 GHz
2 1 3.70 to 4.55 GHz
3 1 4.45 to 5.30 GHz
4 1 5.20 to 6.05 GHz
5 1 5.95 to 6.80 GHz
6 1 6.70 to 7.50 GHz
Band LO multiple (N) AC coupled
MW (Option 513, 526) 0 1 9 kHz to 3.08 GHz
1 2 2.95 to 7.58 GHz
2 2 7.45 to 9.55 GHz
3 2 9.45 to 12.60 GHz
4 2 12.50 to 13.05 GHz
4 4 12.95 to 13.80 GHz
5 4 13.40 to 15.55 GHz
6 4 15.45 to 19.35 GHz
7 4 19.25 to 21.05 GHz
8 4 20.95 to 22.85 GHz
9 4 22.75 to 24.25 GHz
10 4 24.15 to 26.55 GHz
Frequency reference
Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration
accuracy]
Aging rate Option PFR
± 1 x 10
–7
/ year
± 1.5 x 10
–7
/ 2 years
Standard
± 1 x 10
–6
/ year
Temperature stability
20 to 30 °C
Full temperature range
Option PFR
± 1.5 x 10
–8
± 5 x 10
–8
Standard
± 2 x 10
–6
± 2 x 10
–6
Achievable initial calibration accuracy Option PFR
± 4 x 10
–8
Standard
± 1.4 x 10
–6
Example frequency reference accuracy
(with Option PFR)
1 year after last adjustment
= ± (1 x 1 x 10
–7
+ 5 x 10
–8
+ 4 x 10
–8
)
= ± 1.9 x 10
–7
Residual FM
Option PFR
Standard
≤ 0.25 Hz p-p in 20 ms nominal
≤ 10 Hz p-p in 20 ms nominal
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal
resolution 1)
1. Horizontal resolution is span/(sweep points – 1).