For 100A : 1A and 1000A : 1A AC Current Measurements
CT1000 Current Transformer
Specification
Frequency Range: 45Hz to 1kHz
Current Range: 100:1 ratio 10A to 120A rms
1000:1 ratio 100A to 1200A rms
Maximum Input Current: 1000A continuous
2000A for 1 hour
Accuracy (23°C
±
5°C):
±0.2% of specified ratio
Phase Error (23°C
±
5°C): Better than
±
0.1° at 50Hz
Maximum Burden: 0.15VA for specified accuracy (equivalent to the input of
any Voltech power analyzer plus the test leads provided)
Dielectric Strength: 4kV at 50Hz for 1 minute
Temperature Range: Operating: -10°C to +50°C
Storage: -25°C to +85°C
The Voltech CT1000 is a toroidally wound AC current transformer that extends the range of Voltech
power analyzers such as the PM100, PM300 or PM3000A to 1200A rms. Close- tolerance current
ratios provide a measurement accuracy of better than 0.2%, making the CT1000 suitable for preci-
sion-grade measurements.
The CT1000 is housed in a fully insulated, shock-resistant casing with a core aperture that accom-
modates large-diameter cables or rectangular busbars. To minimize the risk of electrical shock, the
CT1000 has 4mm safety sockets for the leads connecting it to the power analyzer. In addition, a
novel over-voltage protection circuit limits the voltage on the output terminals to 50V peak, if the
transformer’s secondary winding is inadvertently open-circuited.
Voltech
CT1000 Current Transformer
VPN: 86-005/5
©
2003 Voltech Instruments. All rights reserved.