Minolta/Land Cyclops 52 Infrared Thermometer
Inst # 20001784
Description
The Minolta-Land Cyclops 52 is a portable thermometer designed for temperature in the
range 600 to 3000° C.
Temperature is displayed digitally.
Digital output is available over RS-232 communication.
Target is viewed simultaneously with the temperature display.
Wavelength of operation chosen to give maximum freedom from errors due to emissivity.
Emissivity compensation in range 0.1 to 1.0.
Three operational modes – Continuous, Peak, and Valley. In the last two the readings
corresponds to the highest and lowest temperature respectively during the period of
operation.
Response time 30 ms (Display 0.6 s).
Accuracy ±0.5% reading.
Wavelength of operation eliminates the effects of water vapor and carbon dioxide in the
atmosphere and the effects of sunlight on temperature readings.
Three calculating modes – Maximum, Mean, or Minimum of a series of readings.
Operation Notes
During operation, the Cyclops 52 displays the temperature in the digital display inside the
viewfinder. When the trigger is pressed the unit attempts a reading, the display will indicate this
by illuminating.
When the temperature is below 500°C the display will illuminate but will not show a number. If
the temperature is between 500-600°C the number will be displayed and the illumination will
blink rapidly alerting the user that this range falls outside the calibration. Between 600-3000°C
the temperature will be indicated and the illumination will be constant. The LCD value is held
for 30s after measurement.
The Cyclops 52 operation is detailed in the Cyclops 52/53 manual.
Page view 0
1 2 3

Summary of Contents

Page 1

Minolta/Land Cyclops 52 Infrared Thermometer Inst # 20001784 Description • The Minolta-Land Cyclops 52 is a portable thermometer designed for tem

Page 2

RS-232 Communication Notes Measured temperature data is available from the digital output terminal in the form of ASCII code, TTL level RS-232. The c

Page 3

the status will display “Below range/Error”, where as if pyrometer supplies a non-zero value, it will display “Acquired.” Again, please note: the C

Comments to this Manuals

No comments