Extended Specications1586A Super-DAQ Precision Temperature Scanner• Measurethermocouples,PRTs,thermistors,dcvoltage,dccurrent,andresistan
10 Fluke Calibration 1586A Super-DAQ Precision Temperature ScannerThermocouple measurement characteristicsRange Temperature Display Resolution
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 11Range Accuracy Fast Sample Rate T.C./ °C Outside18 °C to 28 °C±1
Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A.Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The NetherlandsFluke Calibration. Precision,
2 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner1586A Super-DAQ at a glanceColordisplaywithchannelstatusindicators.Cha
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 3Twenty2-wireorten3-,or4-wireconnections.Built-incablest
4 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner1586A Super-DAQ general specicationsMains Voltage 100 V Setting 90 V to 110
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 51586-2586 High-Capacity Module specications1586-2588 DAQ-STAQ Mu
6 Fluke Calibration 1586A Super-DAQ Precision Temperature ScannerPRT/RTD resistance accuracy Accuracy is given as % of measurement or ohms, wh
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 7Thermistor resistance accuracy Accuracy is given as ± (% of measu
8 Fluke Calibration 1586A Super-DAQ Precision Temperature ScannerThermistor measurement characteristicsRange Temperature Display Resolution So
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 9Thermocouple temperature accuracy Accuracy specifications apply us
Comments to this Manuals