Test /Process Interface (continued)
Resolution: 0.1 degree
Input linearization via NIST table
Accuracy: ±0.1% of full scale DC ± 0.5° C cold-junction
compensation
Connection: Mini jacks
Digital Inputs (All Pods, unless otherwise indicated)
Maximum Frequency: 20 kHz
Minimum Time Interval (for timed inputs): 1 msec
V
H: 4.5 to 24 VDC
V
L: 0 to 0.6 VDC
R
IN: 6.0 k, approximate
Isolation: Optically isolated to 750 VAC RMS, common-mode
Frequency and quadrature inputs configurable for limit logic
Connection: Five-way binding posts
DC Strain Gauge
1, 2, 4 & 10 mV/V
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz
Connection: Screw Terminals
General Specifications
Dimensions (HWD): 12" x 2.5" x 8" (30 cm x 6.4 cm x 20.3 cm)
Weight: 4.2 lb. (1.9 kg)
Operating Temperature: -10° to 50° C (14° to 122° F)
Storage Temperature: -40° to 85° C (-40° to 185° F)
Humidity: 95%, noncondensing
System Time Clock: Crystal-controlled; 1-second resolution
Display: Contrast-adjustable 320 x 240 LCD color touchscreen with
audible feedback and programmable backlight timeout; available
languages include English, French, Italian, German, Spanish
Memory Card: Compact flash 256M supplied
AC Power Adapter supplied (90-264 VAC, 47-63 Hz)
Supporting easel and carry strap supplied
Optional Accessories
Low Voltage and High Voltage Cable Sets
Soft-Sided Carrying Case
Field-Replaceable Battery Pack
External Battery Charger
Weatherproof Container
Reusable Shipping Container
Lockable Portable Case
USB Flash Card Reader
Current Probes: TR25XX (1-3000A) Clamp-on
RR30XX (30/300/3000A) Flexible
PAD
M
EASURING
MEASURINGPAD
SPECIFICATIONS
Scanning, Monitoring, and Recording
Scan Rate: 10 kHz per analog input; 10 kHz per digital input
16-bit A/D conversion
Overall Accuracy (All Measurements): ±0.1% of full scale or better
Overrange (All Measurements): 10% of full scale
Logging Modes:
Manual: A single “snapshot” record taken via touchscreen button
Timed: Records taken automatically between specified start and
stop times; user-specified rate from one second to 24 hours
Event-Triggered: Records taken while any triggering limit violation
or digital input occurs; user-specified rate from one millisecond to
24 hours, or “burst” speed (as fast as possible); overrides timed
logging, and may include “pretrigger” data, if desired
User-enabled beeper for setpoint violation alert
Test /Process Interface
Up to four external sensor interface pods per MeasuringPAD
unit (ordered separately)
Four DC-coupled differential analog inputs of the same type per
pod (see below)
Two digital inputs per pod (see below)
EEPROM per pod, for ID and storage of calibration values
Analog Inputs
Low Voltage Pod
Operating Range (Differential and Common-Mode): ±0 to
30 VDC or 0 to 24 VAC RMS
Input Impedance: 4 M (differential); 1 M (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz
Connection: Five-way binding posts
High Voltage Pod
Operating Range (Differential and Common-Mode): ±0 to
600 VDC or 0 to 480 VAC RMS
Input Impedance: 40 M (differential); 10 M (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz
Connection: Safety banana jacks
Low Current
Operating Range (Differential): ±0 to 20 mADC or 0 to 20
mAAC RMS
Input Impedance: 250 (differential); 1 M (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz
Connection: Five-way binding posts
Thermocouple
Accepts Types B, E, J, K, N, R, S, and T
A versatile, short-term data acquisition solution
for maintenance, troubleshooting and
energy measurement applications
dranetz-bmi.com
1.800.372.6832 [email protected]
tel 732.287.3680 • fax 732.248.1834 • 1000 New Durham Road • Edison, New Jersey 08818 USA
©2008 Dranetz-BMI. All rights reserved. Printed in the United States. Specifications are subject to change without notice.
Revised 07/07/08
DranetzMPad 7/9/08 3:23 PM Page 1
Page view 0
1 2

Summary of Contents

Page 1 - EASURING

Test /Process Interface (continued)Resolution: 0.1 degreeInput linearization via NIST tableAccuracy: ±0.1% of full scale DC ± 0.5° C cold-junctioncomp

Page 2 - MEASURING

Real-Time Data MonitoringMeasuringPAD provides live readout of up to sixteen differential analog inputs andup to eight logic-level digital inputs. As

Comments to this Manuals

No comments