ACCURACYPRESSURE MEASUREMENT
0 to 110% of Full Scale: ±(0.05% of Reading + Floor Term)
Floor Term: 0.005% of FS or display resolution,
whichever is greater
Vacuum: ±(0.25% of Reading
+ Vacuum Floor Term)
Vacuum Floor Term: 0.004 PSI for 16 and 36 PSI
0.01 PSI for 300 PSI
ACCURACYMILLIAMPERES
Accuracy: ±(0.025% of Reading + 0.001 mA)
Range: 0 to 55 mA
Resolution: 0.001 mA
Connection: 4 mm jacks, 19 mm (0.75") spacing
OPERATING TEMPERATURE
0°C to 50°C (32°F to 122°F)
DISPLAY
Screen: 2 lines of up to 6 digits,
5 mm (0.2") high
Display Rate: 2 readings/second (Adjustable to
1, 3, or 3.7 readings/second)
POWER
1 x 9V: 90 hours, typical
DATA/COMMUNICATION
Digital Interface: micro-USB
Includes all eects of operating temperature, linearity, hysteresis,
repeatability and stability, for one year.
All models indicate vacuum, but vacuum specication applies
to 16 PSI, 36 PSI, and 300 PSI models only.
Not recommended for continuous use at high vacuum.
Includes all eects of operating temperature, linearity, hysteresis,
repeatability and stability, for one year.
mA can be displayed as a percentage, where 0 to 100%
corresponds to either 4 to 20 mA or 10 to 50 mA.
Jacks are compatible with safety sheathed banana plugs.
Non-condensing.
LCD readable in sunlight.
Changes to reading rate requires software: CongM30 is a free
download from crystalengineering.net.
Uses one alkaline 9V battery.
Use of USB interface reduces operating time.
Do not use USB in Hazardous locations.
Free Labview drivers available online.
2 LINE
8 CHARACTER
LCD DISPLAY
SENSOR
DIAPHRAGM
LOCATIONS
DIMENSIONS ARE IN MM (INCHES)
137. 2
(5.40)
69.8
(2.75)
1/8" FNPT
PRESSURE PORTS
PRESSURE PORT VOLUMETRIC DATA:
4mm
SHIELDED
BANANA JACKS
29.2
(1.15)
A
B
LP distance to diaphragm ( A to B ): 22.4 (0.88")
LP volume: 2 ml (0.13 cubic inches)
HP distance to diaphragm ( A to B ): 17.8 (0.70")
HP volume: 1 ml (0.06 cubic inches)
MICRO-USB
CONNECTION
DO NOT EXCEED MAXIMUMS
30 Series PSI Data Sheet
4534 . C
Page 1 of 3
30 Series Calibrator PSI
Page view 0
1 2 3

Summary of Contents

Page 1 - 30 Series Calibrator PSI

ACCURACYPRESSURE MEASUREMENT 0 to 110% of Full Scale: ±(0.05% of Reading + Floor Term) Floor Term: 0.005% of FS or display resolution, whic

Page 2

ENCLOSURE All Models: Extruded aluminum with polyester keypad and labels Model 31≤600 PSI: 342 g (12.0 oz) *Model 31≤600 PSI: 428 g (15.

Page 3

IntrinsicSafetyModel Sensors1st PressureRange P/N2nd PressureRange P/NPressure UnitManifoldIS 3 PSISingle Sensor: 1Standard (omit)Dual Sensor: 3 Stain