Atec Astro-Med-Dash-10 User Manual

Browse online or download User Manual for Equipment Atec Astro-Med-Dash-10. Atec Astro-Med-Dash-10 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 1
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Dash 10 Specifications
Input Type Single-ended isolated
User Input Connector Guarded Binding Posts
Input Range 50 mVfs to 500 Vfs
Operating Modes Peak-to-peak or RMS
Input Coupling DC
Input Impedance 1 Megohm
Input Isolation
1500 VACrms for 60 sec., channel to channel,
channel to chassis
Maximum Input 250 Vrms, 500 VDC, 500 Vpeak
CMR (IMR) Greater than 95 dB @ 60 Hz
Common Mode
Voltage
250 Vrms
Frequency Response -1dB at 12 kHz, -2 dB at 18 kHz, -3 dB at 22 kHz
A/D Sample Rate 200 kHz filter disabled
Filter Type 2 pole 12 dB/octave
Filter Response -3dB at 8.75 Hz
Chart Speed Accuracy ± 2%
Zero Suppression
Range
± 500 V or ± 5V on mV ranges
Zero Suppression
Accuracy
± 2% of zero suppression setting
Zero Position
Accuracy
± 0.2%
Battery Charge Time 19 Hours with power off
Battery Life 30 minutes nominal
Operating
Temperature
0 degrees C to 45 degrees C
Storage Temperature -20 degrees C to 80 degrees C
Operational Relative
Humidity
0% to 95% RH non-condensing
Operational Vibration 2G
Weight 32 lbs. without battery, 36 lbs with battery
Timer Speeds
0.01 to 0.09 (0.01 second increments) 0.1 to 0.9
(0.1 sec. increments); 1.0 to 9.0 (1.0 sec.
i
ncrements); 10.0 to 60.0 (5.0 second increments)
Internal Event
Markers
10 interchannel plus 1 system
Internal Annotation
Buffers
Thirty 128-character buffers, One 28-character
system log buffer, One 128-character on-demand
b
uffer
Host Interfaces RS-232, Xon/Xoff or Hardwire; GPIB standard
Baud Rate 300, 600, 1200, 2400, 4800, 9600, 19200
GPIB Address 1 to 31
Data Capture Memory 6 Megasamples (optional - 10 channels)
Stacked Captures 2-8 captures
Data Capture Archive
to/from Floppy Time
22 KBytes per second
Sample Rates
250 kHz, 125 kHz, 50 kHz, 25 kHz, 10 kHz, 5 kHz,
2,5 kHz, 1 kHz, 500 Hz, 250 Hz, 100 Hz, 50 Hz, 25
H
z, 10 Hz, 5 Hz
Playback
Magnification
x1 - x25 (increments of 1); x1/2 - x1/25 (in steps
of 1/x where x is an integer between 2 and 25)
Cold Start Drift 2.0% in 30 minutes
Intrinsic Noise less than 0.5 mm
Baseline Offset less than 0.2 mm
Calibration Accuracy ± 0.8%
RMS Accuracy
± 5% for ranges less than 10VFS
± 2% for ranges greater than 10VFS
RMS Crest Factor 7% at crest factor of 10
Non-linearity < 0.1%
Page view 0
1

Summary of Contents

Page 1 - Dash 10 Specifications

Dash 10 Specifications Input Type Single-ended isolatedUser Input Connector Guarded Binding PostsInput Range 50 mVfs to 500 VfsOperating Modes Peak-to

Comments to this Manuals

No comments