Agilent TechnologiesAgilent E4412A and E4413A Power SensorOperating and Service Guide
x Regulatory MarkingsThe CE mark shows that the product complies with all the relevant European legal Directives (if accompanied by a year, it signif
xiGeneral Safety InformationThe following general safety precautions must be observed during all phases of operation, service, and repair of this in
xii In This Guide ...1 Operation and Service Guide This chapter contains information about initial inspection, performance tests, specifications, ope
xiiiContentsNotices iiGeneral Warranty iiiWarranty Service iiiLimitation of Warranty ivExclusive Remedies ivTechnology Licenses vSafety Summary viS
xiv Cleaning 21Disassembly Procedure 22Reassembly Procedure 23Agilent Sales and Service Offices 24A AppendixE4412A and E4413A Specifications (Seri
xvList of FiguresFigure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP-E26A, respectively) 2Figure 2 Relative Mode Power Measurem
xvi List of TablesTa b l e 1 E4412A and E4413A Specifications A-5Ta b l e 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) A-9Ta b
1Agilent E4412A and E4413A Power SensorsOperating and Service GuideAgilent Technologies1Operation and Service GuideGeneral Information 3Performance Te
1Introduction2 Agilent E4412A and E4413A Power Sensors Operating and Service GuideFigure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 3General InformationWarrantyThe power sensors are warranted and cert
ii Notices© Agilent Technologies, Inc. 1999, 2009No part of this manual may be reproduced in any form or by any means (including elec-tronic storage
1Introduction4 Agilent E4412A and E4413A Power Sensors Operating and Service GuideThese power sensors measure CW power, displayed on a compatible powe
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 5Ta ble 1 E4412A and E4413A Specifications Limit CommentsFrequency
1Introduction6 Agilent E4412A and E4413A Power Sensors Operating and Service Guidea. Limits are in percent of power in Watts. See Figure 2 for relativ
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 7Figure 2 Relative Mode Power Measurement Linearity with Power Meter
1Introduction8 Agilent E4412A and E4413A Power Sensors Operating and Service GuideEXAMPLE A (Amplifier Gain Measurement):Input power (reference) = –35
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 9Ta ble 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm)a. S
1Introduction10 Agilent E4412A and E4413A Power Sensors Operating and Service GuideTa ble 3 E4413A Calibration Factor Uncertainty at 1 mW (0 dBm)a. S
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 11InstallationInitial InspectionInspect the shipping container for d
1Introduction12 Agilent E4412A and E4413A Power Sensors Operating and Service GuideStorage and ShipmentEnvironmentThe instruments should be stored in
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 13Operation PrecautionsIf the following energy and power levels are
iiiCertificationAgilent Technologies certifies that this product met its publishedspecifications at the time of shipment. Agilent further certifies
1Introduction14 Agilent E4412A and E4413A Power Sensors Operating and Service GuidePower Meter CalibrationsFollow the calibration procedures given in
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 15Performance TestStanding Wave Ratio (SWR) and Reflection Coefficie
1Introduction16 Agilent E4412A and E4413A Power Sensors Operating and Service GuideTa ble 5 Power Sensor SWR and Reflection Coef ficient for the E441
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 17Replaceable PartsTable 6 is a list of replaceable parts. Figure 4
1Introduction18 Agilent E4412A and E4413A Power Sensors Operating and Service GuideFigure 4 Illustrated Parts Break down
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 19Ta ble 6 Replaceable PartsReference DesignationPart Number Qty D
1Introduction20 Agilent E4412A and E4413A Power Sensors Operating and Service GuideServiceService instructions consist of principles of operation, tro
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 21If error message 241 or 310 is indicated on the power meter, suspe
1Introduction22 Agilent E4412A and E4413A Power Sensors Operating and Service GuideClean the connector face using a cotton swab dipped in isopropyl al
Introduction 1Agilent E4412A and E4413A Power Sensors Operating and Service Guide 23Reassembly Procedure1 Replace the magnetic shields and the plastic
iv Limitation of WarrantyThe foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supp
1Introduction24 Agilent E4412A and E4413A Power Sensors Operating and Service GuideAgilent Sales and Service OfficesTa ble 7 Agilent Sales and Servic
A-1Agilent E4412A and E4413A Power SensorsOperating and Service GuideAgilent TechnologiesAAppendixE4412A and E4413A Specifications (Serial Prefixes Be
AAppendixA-2 Agilent E4412A and E4413A Power Sensors Operating and Service GuideTa ble A-1 E4412A and E4413A Specifications (Serial Prefixes Below U
Appendix AAgilent E4412A and E4413A Power Sensors Operating and Service Guide A-3a. Limits are in percent of power in Watts. See Figure 2 for relativ
AAppendixA-4 Agilent E4412A and E4413A Power Sensors Operating and Service GuideTa bl e A-2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (S
Appendix AAgilent E4412A and E4413A Power Sensors Operating and Service Guide A-5Ta bl e A-3 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (
AAppendixA-6 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
© Agilent Technologies, Inc. 1999, 2008, 2009Printed in MalaysiaMay 29, 2009E4412-90013www.agilent.comAgilent TechnologiesContact usTo obtain service,
vRestricted Rights LegendThe Software and Documentation have been developed entirely at privateexpense. They are delivered and licensed as “commerci
vi Safety SummaryThe following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with
viiSafety SymbolsThe following symbol on the instrument and in the documentation indicates precautions that must be taken to maintain safe operation
viii PROTECTIVE CONDUCTOR TERMINAL.Frame or chasis TERMINAL.Equipotentiality.On (Supply).Off (Supply).Equipment protected throughout by DOUBLE INSULA
ixOut position of bi-stable push control.