Atec Agilent-6030A User Manual

Browse online or download User Manual for Power suppliers Atec Agilent-6030A. Atec Agilent-6030A User Manual [es]

  • Download
  • Add to my manuals
  • Print
  • Page
    / 4
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
This series of 200 watt and 1000
watt DC power supplies take the
place of multiple power supplies
in your test system by providing
maximum power at a variety of
operating points.
Industry standard SCPI commands
and VXIPlug&Play drivers make
system integration easy. Using the
serial link, up to 16 power supplies
can be connected through one
GPIB address. These power supplies
have excellent electrical efficiency,
making them a good choice for
large systems.
Application Notes:
10 Hints for Using Your Power Supply
to Decrease Test Time
5968-6359E
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
Agilent DC Power Supplies
for Base Station Testing
5988-2386EN
Basic DC Power Supplies essential features for a tight budget
Specifications
(at 0
˚
to 55
˚
C unless
otherwise specified)
6030A 6031A 6032A 6033A 6035A 6038A
More detailed specifications at
www.agilent.com/find/6030
25
Autoranging outputs provide maximum power at a variety
of operating voltages
Analog/resistance control of output voltage and current
Series and auto-parallel connections of multiple supplies
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
6030A, 6031A, 6032A, 6033A, 6035A, 6038A
Single-Output, Autoranging
200 W and 1000 W GPIB
Number of outputs 111111
GPIB Yes Yes Yes Yes Yes Yes
Output ratings
Output Voltage 0 to 200 V 0 to 20 V 0 to 60 V 0 to 20 V 0 to 500 V 0 to 60 V
Output Current 0 to 17 A 0 to 120 A 0 to 50 A 0 to 30 A 0 to 5 A 0 to 10 A
Maximum power watts 1,200 W 1,064 W 1,200 W 242 W 1,050 W 240 W
Autoranging output
V1, I
1
200 V, 5 A 20 V, 50 A 60 V, 17.5 A 20 V, 10 A 500 V, 2 A 60 V, 3.3 A
V2, I
2
120 V, 10 A 14 V, 76 A 40 V, 30 A 14 V, 17.2 A 350 V, 3 A 40 V, 6 A
V3, I
3
60 V, 17 A 7 V, 120 A 20 V, 50 A 6.7 V, 30 A 200 V, 5 A 20 V, 10 A
Programming accuracy Voltage 0.035% 0.035% 0.035% 0.035% 0.25% 0.035%
at 25°C ±5°C +145 mV +15 mV +40 mV +9 mV +400 mV +40 mV
Current 0.2% 0.25% 0.2% 0.15% 0.3% 0.09%
+25 mA +250 mA +85 mA +20 mA +63 mA +10 mA
Ripple and noise Voltage rms 22 mV 8 mV 8 mV 3 mV 50 mV 3 mV
20 Hz to 20 MHz p-p 50 mV 50 mV 40 mV 30 mV 160 mV 30 mV
Current rms 15 mA 120 mA 25 mA 30 mA 50 mA 5 mA
Readback accuracy Voltage 0.08% 0.08% 0.08% 0.07% 0.5% 0.07%
at 25°C ±5°C +80 mV +7 mV +20 mV +6 mV +200 mV +50 mV
Current 0.36% 0.4% 0.36% 0.3% 0.5% 0.2%
+15 mA +100 mA +35 mA +25 mA +50 mA +11 mA
Load regulation
Voltage 0.01%+ 5 mV 3 mV 5 mV 2 mV 40 mV 3 mV
Current 0.01%+ 0.01%+ 0.01%+ 0.01%+ 0.03%+ 0.01%+
10 mA 15 mA 10 mA 9 mA 34 mA 5 mA
Line regulation
Voltage 0.01%+ 0.01%+ 0.01%+ 0.01%+ 0.03%+ 0.01%+
5 mV 2 mV 3 mV 1 mV 17 mV 2 mV
Current 0.01%+ 0.01%+ 0.01%+ 0.01%+ 0.03%+ 0.01%+
5 mA 25 mA 10 mA 6 mA 17 mA 2 mA
Transient response time Time 2 ms 2 ms 2 ms 1 ms 5 ms 1 ms
10% step change Level 150 mV 100 mV 100 mV 50 mV 200 mV 7 5 mV
V1
I1 I2 I3
Volts
P1
P2
P3
Amps
V2
V3
Autoranging Output:
Page view 0
1 2 3 4

Summary of Contents

Page 1 - 200 W and 1000 W GPIB

This series of 200 watt and 1000watt DC power supplies take theplace of multiple power supplies in your test system by providingmaximum power at a var

Page 2 - (Continued)

Size0.6128mm5.04"12.7mm0.5"425mm16.75"132.6mm5.2"12.7mm0.5"421.6mm16.6"77mm3.03"2.5mm0.1"RearSideTo pTerminal

Page 3

27Ordering InformationOpt 001 Front panel has only line switch, line indicator, and OVP adjust (6030A–33A and 6038A only)Opt 100 87 to 106 Vac, 48 to

Page 4 - Agilent Technologies

Your Requested Excerpt from the Agilent System and Bench Instruments Catalog 2006The preceding page(s) are an excerpt from the 2006 System and Bench I

Comments to this Manuals